Sheng Taike intelligence measuring appliance
configuration software user's manual Welcome your excellency to select Sheng Taike the
intelligence measuring appliance and Sheng Taike intelligence
measuring appliance configuration software, Sheng Taike intelligence
measuring appliance configuration software is specially aims at Sheng
Taike the intelligence measuring appliance to design, has the scene
configuration, active and the passive adjustment, temperature
compensation, the linearity compensates, the permanent current source
output and so on the many kinds of functions. You in use Sheng
Taike the intelligence measuring appliance and Sheng Taike in front of
intelligence measuring appliance configuration software, please by all
means must carefully reads the textbook user's manual, in order to
avoid creates the nonessential trouble! First, system demand 1) operating system: Windows
95/98/Me/2000/XP or above 2) CPU:350MHZ or is quicker 3) memory: Windows 95/98/ME: 64MB or is
bigger, Windows 2000/XP: 128MB 4) port: (When RS232 port or USB port with USB
port, must add seperately USB - RS232 to switch over electric cable) 5) corresponds Modem (not to be possible to use the
market RS232 electric cable to replace) Second, software installment Intelligence measuring appliance configuration
software installs the compact disc in Sheng Taike to put in the light
to drive, turns on the HTSoft/Debug folder, found the Setup.exe
document, double-clicks the movement, single-clicks [ Next ] then to
be possible to carry on the next step of installment, after firmly
arranges installs the way above, single-clicks [ Next ] the button
then to be possible to carry on the next step, single-clicks [ Next ]
to determine the installment, the installment finished, now may
discover the installation procedure already automatically to lay aside
on the tabletop quickly to attempt the sign: HTSoft,
double-clicks it then to be possible to move Sheng Taike intelligence
measuring appliance configuration software Third, software operation Operation flow After the user attains brand-new changes the
delivering intelligence board together, may operate the flow according
to below to carry on the operation: Starts to move | Output current calibration | Measuring range establishment | Measuring range migration | Establishment constant pressure source/permanent current
source | The gain establishment (according to needs to be
possible to elect) | Input pressure calibration | Other parameter establishments: The damping time,
a zero adjustment, the demonstration pattern, the pressure unit,
report to the police the electric current, the filter constant,
(according to needs to be possible to elect) | The linearity compensates (according to needs to be
possible to elect) | Temperature compensation (according to needs to be
possible to elect) | Performance examination | Finished Starts to move Puts through the intelligence to change the
delivering power source, and connects the good computer with to change
delivering correspondence Modem, moves Sheng Taike intelligence
measuring appliance configuration software, if all is normal, after
enters the main contact surface, the procedure bulletin considers
discovers Sheng Taike the intelligence measuring appliance, otherwise,
can report when the error message, appears the error message,
single-clicks < tries the button again >, if still reported the error
message, please see also the question as soon as to carry on the
solution. After discovered the product, single-clicks < determination >
the button to enter the host contact surface, now, then might carry on
changes the delivering parameter establishment or other correlations
operates. Output current calibration Sheng Taike the intelligence changes delivering or
changes when the delivering multi-purpose board leaves the plant, its
output current all passes through the calibration, generally speaking,
does not need the user to carry on the output current calibration
operation again. If the user discovered the intelligence changes
delivering or changes on the delivering multi-purpose board to output
when the actual magnitude of current and the theory value deviation is
bigger, moreover this kind of deviation is because changes the
delivering output current calibration not to cause, then needs to
carry on the output current calibration. If the user is unable to judge this kind of deviation is causes
by the sensor non-linear error, or because the output magnitude of
current calibration has not caused, may refer to the frequently asked
questions faq to carry on the judgement. Single-clicks [ debugging/output current calibration ] the menu,
appears the electric current calibration contact surface, and can
automatically read out changes the delivering electric current output
calibration value, the 4mA electric current calibration value
generally in "2,400" (hexadecimal system value) about, 20mA electric
current calibration value generally in "C400" (hexadecimal system
value) about. 4mA output current calibration: Will change on the
delivering ring circuit to serially connect an ampere meter, and will
single-click in the first frame [ transmission ] the button,
simultaneously will observe on the ampere meter the magnitude of
current, will carry on according to the ampere meter on value to the
4mA output current calibration value correspondingly increases or
reduces, until the ampere meter on value demonstration will be
4.000mA. By now, then might the button preservation data to
change delivering according to [ revision ]. Calibrates
successfully, the procedure can report revises the successful
information. 20mA output current calibration: Single-clicks in the
second frame [ transmission ] the button, simultaneously observes on
the ampere meter the magnitude of current, the value calibrates the
value according to the ampere meter on to the second frame in to carry
on correspondingly increases or reduces, until the ampere meter on the
value demonstration is 20.000mA. By now, then might the button
preservation data to change delivering according to [ revision ].
Calibrates successfully, the procedure can report revises the
successful information. In the third frame value positioning is establishes for the
convenience user adjustment calibration value. After the output current calibration finished, may inspect the
electric current output in often the value electric current output
linear. After calibrates successfully, the closure current contact
surface, changes delivering to be able to carry on the normal pressure
gathering work, does not need reto start. Measuring range establishment The measuring range establishment uses in to
establish the sensor the measuring range, zero pressure and the full
spot pressure generally are refers to the sensor as follows the
measuring range, the operation: 1st, establishment zero pressure: Single-clicks [ establishment/measuring range to demarcate/a
zero pressure ] the menu, newly appears the window can automatically
read out currently changes the delivering zero pressure value, inputs
the new value in a change zero pressure value fence, clicks on [
revision ] the button, then completes the operation, the full spot
pressure establishment operation same, the here no longer states. After the software operation/basic function/measuring range
migration changes delivering to leave the plant, if the user wants to
carry on the measuring range migration, may through change 4mA or the
20mA pressure point of force realizes. Single-clicks [ establishment/measuring range to demarcate the
/4mA pressure ] the menu, newly appears the window can automatically
read out currently changes the delivering 4mA pressure value in to
change in a 4mA pressure value fence to input the new value, clicks on
[ revision ] the button, then completes the operation, the 20mA
pressure establishment operation same. If in already changed delivering after the input pressure
calibration to carry on the measuring range migration again, generally
speaking also needed reto carry on the input pressure calibration, in
some situations was allowed not to need after the measuring range
migration to input the pressure calibration again, namely the passive
measuring range migration calibration, the special details saw also
the frequently asked questions faq. Establishment constant pressure source/permanent current
source Sheng Taike the intelligence changes delivering in
to set at the programmable constant pressure source and the
programmable permanent current source, uses in to the sensor power
supply, single-clicks [ establishment/constant pressure
source/permanent current source ] the menu, newly appears the window
can automatically read out currently changes the delivering constant
pressure source/permanent current source value. Inputs the new
value in a change constant pressure source/permanent current source
value fence, the value in (0-255) between, clicks on [ revision ] the
button, then completes the operation. Explained: 1) when the value chooses 255, then closure
constant pressure source and permanent current source; 2) if uses the constant pressure source hardware line,
sensor power line voltage V = constant pressure source value *5/255
(V); 3) if uses the permanent current source hardware line,
sensor power supply electric current I = (1- permanent current source
value /,255) *5/150 mA (When I*Rsense<5V, I for has flowed sensor
electric current, Rsense is sensor 内阻.) Establishment gain The gain establishment is helpful to the
enhancement changes the delivering sampling resolution, in has the
measuring range to migrate or other small signal sensors (for example:
The proliferation silicon sensor, the ceramics press the anti-
sensor and so on) specially useful, may change the CPU interior
enlargement factor through the gain establishment, causes the data
scope which may gather to be bigger, operation as follows:
Single-clicks [ debugging/gain establishment ] the menu, in
appears in the window chooses the good sensor, and inputs the sensor a
zero pressure and the full spot pressure, after as well as the
measuring range migration 4mA pressure and the 20mA pressure, then
click on [ ==> ] the button, after the system can the automatic
computation migration enlargement factor, the migration ratio, and CPU
interior some parameters, then clicks on [ transmission data ] the
button then. Input pressure calibration For changes delivering to exert a zero pressure,
when on changes when the delivering output 4mA electric current
corresponds the pressure, single-clicks [ debugging/input pressure
calibration ] the menu, appears the input pressure calibration contact
surface, single-clicks the calibration zero spot, changes delivering
then to be able automatically to preserve the current spot is the zero
spot, simultaneously changes the delivering output 4mA electric
current, and appears the report to operate the successful contact
surface. Again gives changes delivering to exert the full spot pressure,
single-clicks calibrates the full spot, changes delivering then to be
able automatically to preserve the current spot is the full spot,
simultaneously changes the delivering output 20mA electric current,
and appears the report to operate the successful contact surface. Establishment damping time The damping time is helpful to suppresses the
scene signal the undulation, the damping time in a big way, changes
the delivering response to be slower, its establishment range of
validity between 0.1s-99.9s. Establishment operation as follows:
Single-clicks [ establishment/damping time ] the menu, newly
appears the window can automatically read out currently changes the
delivering damping time value. Inputs the new damping time value in a change damping time
fence, clicks on [ revision ] the button, then completes the
operation. Establishment zero adjustment Changes delivering when the installment as a
result of 偏斜 or the pipeline other additional pressures, when
creates a zero pressure, its output certainly is not 4mA, in order to
readjusts zero, may press a key the way carries on, also may carry on
through the software establishment zero adjustment way adjusts zero as
follows, the establishment zero adjustment operation:
Demonstrated the pattern changes the demonstration pressure
pattern, lets change delivering to be at under the zero pressure
condition, looked on its screen demonstrated the pressure value,
nominal zero pressure value reducing place this value will be the zero
adjustment value, the establishment as follows: Single-clicks [
an establishment/zero adjustment ] the menu, newly appears the window
can automatically read out currently changes the delivering zero
adjustment value. Inputs the new zero adjustment value in a change zero adjustment
value fence, clicks on [ revision ] the button, then completes the
operation. The establishment demonstrates the pattern The intelligence changes delivering to be allowed
to dispose the LCD display monitor, with changes the delivering
interior by the real time display the data, matches by the pressed
key, but also may demonstrate each kind of pattern menu. Under
the normal survey condition, the LCD display monitor may demonstrate
the current pressure, changes the delivering output current, or
percentage three kind of center one kind. The change
demonstrates the pattern operation as follows: Single-clicks [
establishment/demonstration pattern ] the menu, newly appears the
window can automatically read out currently changes the delivering
demonstration mold to show. Chooses the new demonstration
pattern in a change demonstration pattern fence, clicks on [ revision
] the button, then completes the operation. Establishment pressure unit When the intelligence changes the delivering
disposition LCD display monitor, in the demonstration pattern for the
demonstration pressure condition under, sometimes needs to change the
demonstration the pressure unit, its operation as follows:
Single-clicks [ establishment/pressure unit ] the menu, newly
appears the window can automatically read out currently changes the
delivering pressure unit. Chooses the new pressure unit in a
change pressure unit fence, clicks on [ revision ] the button, then
completes the operation. Explained: The change pressure unit, to changes the
delivering survey and the output cannot have any influence, only plays
the change demonstration role. The establishment reports to the police the electric
current Changes delivering to be at the work in the
abnormal state time, for example: Crosses the pressure, the
sensor damage, the outside violent qrm and so on, by now changed
delivering to be able auto-alarm, and the passable establishment has
reported to the police the electric current caused to change
delivering when reporting to the police the output current was 3.6mA
or 22.8mA, the operation as follows: Single-clicks [
establishment/to report to the police electric current ] the menu,
newly appears the window can automatically read out current changes
delivering reports to the police the magnitude of current.
Reports to the police in a magnitude of current fence in the
change to choose wants to report to the police the magnitude of
current, clicks on [ revision ] the button, then completes the
operation. Establishment filter constant The filter constant may eliminate changes the
delivering finally several characters beating, the value in a big way,
stability better, but changes delivering to be keen to mentions in
disguised form badly; Single-clicks [ establishment/filter
constant ] the menu, newly appears the window can automatically read
out currently changes the delivering filter constant value, inputs the
new value in a change filter constant value fence, clicks on [
revision ] the button, then completes the operation. Synthesis establishment Front introduced each parameter establishment
also may disposable read in, speeds up the production speed,
single-clicks [ establishment/product parameter establishment ] the
menu, newly appears in the window the value to tacitly approve the
parameter for the product, the user may according to the need, revises
some or many parameters, also may click [ reads fetching according to
] the button, the read-out changes the delivering interior parameter,
the inspection data accuracy. After the parameter revision
finished, clicks on [ revision parameter ] the button, then may
disposable pass on the parameter to changes delivering. Often value electric current output Single-clicks [ tool/often value electric current
output ] the menu, can appear a new window, in the window may the
direct selection output 3.6mA,4mA,8mA, 12mA, 16mA, 20mA, 22.8mA, also
may output in 3.6mA to the 22.8mA scope in free magnitude of current The attention, after changes the delivering output
often the value electric current, output current not pressure control.
Withdraws from this condition, after namely the closure window,
will change delivering the again gauge pressure. After chooses
the magnitude of current, single-clicks [ determination ] the button,
changes delivering is the output choice magnitude of current.
Also may first select other magnitudes of current, then in
inputs the magnitude of current in a right side fence (for example:
15.689mA), clicks on [ determination ] the button, changes
delivering namely to output 15.689mA On-line test The on-line test may use in to examine changes
delivering whether has connected with the computer, operation as
follows: Single-clicks [ debugging/on-line test ] the menu, can
appear the on-line window, and appears the on-line information, after
the on-line success, in the window can appear changes the delivering
interior the related parameter. Chooses the port The system tacitly approves the port is the com1
port, if in the user computer does not have in the com port, or the
com1 port has been taken by other equipment, but also has other com
port to be possible to use, through chooses the port to change tacitly
approves the port. Single-clicks [ debugging/to choose port ]
the menu, appears the choice port window, after chooses the good port,
clicks [ turns on ] the button then. If chooses the port does
not exist, or the choice port has been taken by other ports, then can
appear the error message. If in the user computer does not have
the com port, but has in the USB port, may again buy USB-RS232 in the
market to switch over the electric cable, the USB string mouth
transformation is the COM port, again in chooses in the port to choose
the corresponding COM port then to be possible. The linearity compensates The software changed delivering above V3.06
edition intelligence all to provide five lines between two points
compensatory function, and the linearity compensated the points to be
allowed to choose at will in 1-5 before, the linearity compensated,
the request first carried on the input pressure calibration, after the
input pressure calibration finished, examined its linear, if linear
did not conform to the requirement, might carry on the linearity to
compensate, has rests the operation as follows: Single-clicks [
debugging/linearity to compensate ] the menu, appears the linearity to
compensate the contact surface, here linearity compensated the spot to
be able to select second, third, or fourth (first and fifth already
has calibrated), first input the zero pressure value and the full spot
pressure value, the system can tacitly approve according to 25%FS,
50%FS, 75%FS three carries on compensates, the user also might
voluntarily adopt other to press the point of force to carry on
compensates. When compensates according to the %FS way input,
may not use the input pressure value. Carries on when the linearity compensates, first clicks [ starts
] the button, the system starts the automated analysis interior
parameter, after the analysis finished, for changes the pressure which
delivering exerts second assigns, clicks on [ gathering ] the button,
the system read-out linearity compensates the value, then clicks on [
transmission data ] the button, then the second line between two
points compensates finished, third, the fourth line between two points
compensates with second. Attention: Second, third, the
fourth pressure size certainly needs to defer to the from infancy to
maturity arranged in order. If among only wants to compensate two, may cancel fourth which
selects, only compensates third, second then; If among only wants to compensate, may cancel third, fourth
which selects, only compensates second then. Temperature compensation The software changed delivering above V3.05
edition intelligence all to provide ten temperature compensations
function, and the temperature compensation points were allowed to
choose at will in 1-10, when chose 1 o'clock, namely established for
does not bring the temperature compensation function, had rests the
operation as follows: Single-clicks [ debugging/temperature
compensation ] the menu, first chooses the temperature correction
points in a temperature compensation fence, after chooses, clicks on [
transmission points ] the button. Again will change delivering
to set warm to controls in the box, (warm will go past high regarding
the PTC characteristic temperature sensor, temperature output signal
bigger), when temperature compensation should the way which gradually
will elevate according to the temperature carry on compensates (namely
first to compensate a temperature lowly, finally as soon as will
compensate a temperature to be highest); (Warm goes past high
regarding the NTC characteristic temperature sensor, temperature
output signal smaller), when temperature compensation should the way
which gradually reduces according to the temperature carry on
compensates (namely first to compensate a temperature high, finally as
soon as compensates a temperature to be lowest); Whenever a temperature compensation spot, best pauses 30
minutes, after treats changes delivering interior temperature complete
equilibrium, after chooses well 校准? clicks on [ calibration ]
the button then. When calibrates first, can appear "must transmit lacks the
province value?" The prompt, clicks [ is (Y) ], then transmits
lacks the province data, also will be second later all compensates all
to press tacitly approves the value to transmit to changes delivering
inside. If is first time compensates, then clicks [ is (Y) ], if will be
later carries on the renewal to compensate again, revision simple
point data, then will click on [ otherwise (N) ], in order to avoid
will flush out other to compensate the points according to. In the temperature compensation process, simultaneously needs to
choose two carries on the temperature to demarcate, like this may let
change the delivering correct demonstration temperature function.
Demarcates the process to carry on in the first fence, first
inputs in the first fence corresponding column demarcates two
temperatures values (namely a zero temperature and full spot
temperature), after the temperature arrives corresponding temperature,
first chooses the corresponding temperature scale fixed point in the
first fence, clicks on [ calibration ] to be possible, then similarly
carries on the second calibration again. Data restoration The software changed delivering above V3.02
edition intelligence all to increase the data to restore the function,
if user in scene adjustment parameter time, has messed up the internal
parameter, will be allowed to restore the function through the data to
change delivering to restore to leaves the plant the establishment,
the operation as follows: Restores before the data, certainly
must first to change the delivering examination production date
through software, if production date unknown, then cannot carry on the
data restoration! After the read-out production date, single-clicks [ tool/data to
restore ] the menu, then may carry on the data restoration. The examination changes the delivering production
information Single-clicks [ examination/to change delivering
software edition ] the menu, may read out currently changes the
delivering software edition. Single-clicks [ examination/production date ] the menu, may read
out currently changes the delivering production date (to change
delivering to the V3.04 edition above to be effective). Single-clicks [ examination/product serial number ] the menu,
may read out currently changes the delivering product serial number
(to change delivering to the V3.04 edition above to be effective). |