> Abstract: This article to the water
quality automatic monitor system technical constitution and the
technical key, as well as the on-line water quality autoanalyzer
technical principle, sent the people present situation and the
existence question has carried on the narration. Discussed the
conventional five parameters, the chemical oxygen demand with emphasis
(COD), the permanganate index, 总有机碳 (TOC), the ammonia
nitrogen, the total phosphorus and the UV on-line autoanalyzer. Key word: Water quality automatic monitor system
autoanalyzer 1 foreword The water quality on-line automatic monitor system
is a set take the on-line automated analysis instrument as the core,
the utilization modern sensor technology, the automatic survey
technology, the automatic control technology, the computer application
technology as well as correlation special-purpose analysis software
and the communication network composes a comprehensive on-line
automatic monitor data, statistical, processing monitor data, but
printout date, week, month, season, year even uniform data as well as
date, week, month, season, year maximum value, minimum value and so on
each kind of monitor, statistical report and graph (clavate chart,
diagram of curves multi- trajectory diagrams, contrast chart and so
on), And may input the central database or access the net.
The collection and may the long-term memory assign the monitor
data and each kind of movement material, the environment material
prepare the retrieval. The system has the monitor project to
exceed the allowed figure and the sub- station condition signal
demonstrated, reports to the police the function; The automatic
movement, the power cut protection, comes the electricity
self-recovery function; The maintenance examines the oak
condition test, is advantageous for function and so on routine
maintenance and emergency breakdown processing. The implementation water quality automatic monitor, may realize
the water quality real-time continuous monitoring and the
long-distance monitoring, achieved and grasps the main basin key cross
section water body the water quality condition, the early warning
forecast is significant or the basin water pollution accident, solves
the cross administrative division water pollution accident dispute,
inspector general total quantity control system goal and so on
realization situation, platoon unconventional sign situation. 2 water quality automatic monitors technology 2.1 water quality automatic monitors system
constitution In the water quality automatic monitor system network,
the central station dials through the satellite and the telephone
holds in the mouth two kind of communications ways realization to each
sub- station real-time surveillance, the long-distance control and the
data transmission function, is entrusted with stands also may through
the telephone digit dialing way realization to holds 管子?the
real-time surveillance, the long-distance control and the data
transmission function, other have produced presently the real-time
surveillance and the data transmission after the authorized related
department passable telephone digit dialing way which stands to the
climax or energy. Each stature station is an independent integrity water quality
automatic monitor system, generally constitutes by 6 major subsystems,
including: The sampling system, the pretreatment system, the
monitor instrument system, the PLC control system, the data
acquisition, processing and the transmission subsystem and the long
range data administrative center, the monitor 站房or monitors the
hut. At present, in the water quality automatic monitor system
subsystem and the long range data administrative center, the monitor
站房or monitors the hut. At present, in the water quality
automatic monitor system sub- station constitution way approximately
has three kinds: (1) by or a many small multi- parameters water quality
autoanalyzer (for example: YS1 Corporation and the HYDROLAB
Corporation's conventional five parameters analyzer) composes the sub-
station (many combinations may use in to survey different water depth
water quality). Its characteristic is the instrument may
directly put Yu Shuizhong to survey, the system constitution nimble is
convenient. (2) fixed formula station: For compares traditional the
system composition way. Its characteristic is monitors the
project the choice scope to be wide. (3) mobile formula station: One kind completely installs
for the fixed formula station instrumentation equipment to a trailer
(monitors hut) on, may according to need to migrate the place, also
may think is half fixed formula station. Its characteristic is
composes the cost to be higher. Various units through the water sample transportation circuitry,
the signalling system, the compressed air transportation circuitry,
the pure water transportation circuitry realization mutually relate. May * the natural very high water quality automatic monitor
system, have simultaneously to have 4 essential factors, namely (1)
high grade system equipment; (2) complete system design;
(3) strict construct control; (4) responsible movement
management. 2.2 water quality automatic monitors technical key 2.2.1 picks the water unit Including the water pump, the pipeline, supply power and
installs the structure to be partial. In the design must to in
each kind of climate, the terrain, the water level change and the
water the silt and so on proposed the corresponding solution measure,
can be automatic continuously with the overall system synchronization
work, provides to the system may *, the effective water sample. 2.2.2 matches the water unit Including water sample pretreatment installment,
self-cleaning installment and auxiliary body. Matches the water
unit directly to the automatic monitor instrument water supply, has
on-line eliminates the silt and on-line filters, manual and the
automatic pipeline counter- flushing and eliminates the algae
installment; Its water quality, the hydraulic pressure and the
water volume should satisfy automatically monitors the instrument the
need. 2.2.3 analyses units Are composed by a series of water quality automated
analysis and the metering equipment, including: The water
temperature, PH, 溶解?(DO), the conductivity, the turbidity, the
ammonia nitrogen, the chemical oxygen demand, the permanganate index,
总有机碳 (TOC), the total nitrogen, the total phosphorus, the
nitrate, the phosphate, the cyanide, the fluoride, the chloride, the
phenol class, the oil class, the metallic ion, 水位? the current
capacity/current capacity/flows to takes into account composition and
so on the automatic sampling. Each main on-line automated
analysis instrument development present situation 3rd will relate in
detail. 2.2.4 control units Including: (1) systems control cabinet and systems
control software; (2) data acquisition, processing and memory
and its application software; (3) wire communication and
satellite communication equipment. 2.2.5 sub- stations 站房and necessary facility Including: (1) 站房main body; (2)
necessary facility 3 on-line automated analyses instruments
development 3.1 outlines The water quality automatically monitored the instrument
still during the development, country and so on Europe, America,
Japan, Australia had some specialized merchant productions. At
present, includes after the maturer conventional project: Water
temperature, PH, 溶解?(DO), conductivity, turbidity, redox
potential (ORP), speed of flow and water level and so on. The
commonly used monitor project includes: COD, permanganate index,
TOC, ammonia nitrogen, total nitrogen, total phosphorus. Other
also have: Fluoride, chloride, nitrate, nitrite, cyanide,
sulfate, phosphate, active chlorine, TOD, BOD, UV, oil class, phenol,
chlorophyll, metallic ion (for example six prices chromium) and so on. The present autoanalyzer generally has the following function:
The automatic measuring range transforms, the remote control,
the standard output connection and the numeral demonstrated,
self-cleaning (when clean has data locking function), the condition
self-checks and reports to the police the function (for example:
The liquid divulges, line clogging, the superhigh measuring
range, the instrument internal hyperpyrexia, the reagent with the
ruler, the high/low density, the power failure and so on), does the
revolution and the power failure protection, comes the electricity
self-recovery, instrument and so on COD, ammonia nitrogen, TOC, total
phosphorus, total nitrogen to have automatically demarcates the
adjustment function. 3.2 conventional five parameters analyzers The conventional five parameters analyzers frequently
use flow 通式 the multi- sensors survey pond structure, does not
have a zero drifting, does not need the baseline to adjust, has the
integrated biology clean and the compressed air purifier. For
example: English ABB Corporation produces the EIL7976 multi-
parameters analyzer, French Polymetron Corporation produces the
conventional five parameters analyzers, Australian GREENSPAN
Corporation produces Aqualab multi- parameters analyzer (including
conventional five parameters, ammonia nitrogen, phosphate).
Another kind of type ("4+1") the conventional five parameters
autoanalyzers representative is the MP2000 multi- parameters on-line
not nature analyzer which French SERES Corporation produces, its
characteristic is the instrument structure is compact. The conventional five parameters survey principles respectively
are: The water temperature for the temperature sensor law
(Platinum RTD), PH is the glass or 锑电?the law, DO is the golden
- silver membrane electrode law (Galvanic), the conductivity for the
electrode law (exchange 阻抗?, the turbidity for the light
research method (transmission principle or infrared scattering
principle). 3.3 chemical oxygen demands (COD) analyzer The COD on-line autoanalyzer main technical principle
has six kinds: (1) the potassium bichromate dispels - the
luminosity to measure the mensuration; (2) the potassium
bichromate dispels - the coulomb titrimetric method; (3) the
potassium bichromate dispels - the reduction method; (4) UV
counts (254nm); (5) 氢氧?and the ozone (mixes oxidant) to
oxidize - the electrochemistry to measure the mensuration; (6)
the ozone oxidation - electrochemistry measures the mensuration. Says from the principle, the method (3) approaches the GB code
method, the method (2) also is the recommendation unified method.
The method (1) already used on the fast COD determination
instrument. Method (5) and method (6) although does not belong
to GB code or the recommendation method, but in view of the fact that
its has the movement may * and so on the characteristic, in the
practical application, only have carries on its analysis result and
the GB code method compared to after experiments and carries on the
suitable adjustment, then gives to approve. But the method (4)
uses in table piece water quality COD, although obtained a more
widespread application in Japan, but the European and American various
countries not yet applies (has not obtained administrative hosts and
visitors department approval), still had to carry out the related
research in our country. Says from the analysis performance, on-line COD meter survey
scope generally in 10 (or 30) ~2000mg/l, therefore, the present
on-line COD meter only can satisfy the source of pollution on-line
automatic monitor the need, should use in with difficulty the surface
water automatic monitor. Moreover, with uses the
electrochemistry principle the instrument to compare, uses dispels -
the reduction method, dispels - luminosity method instrument minute
cyclical generally longer somewhat (10min~2h), former generally is
2~8min. Says from the instrument structure, uses the on-line COD meter
which the electrochemistry principle or UV counts generally compared
to use dispels - the reduction method, dispels - the luminosity method
instrument structure simply, and because former enters the type and
the reagent easily joins the system (to pump, tube few), therefore not
only in operation more convenient, moreover its movement may * the
nature be also better. Says from the maintenance difficulty degree, because dispels -
the reduction method, dispels the reagent type which - the luminosity
law uses many, pump pipe system more complex, therefore in reagent
replacement as well as pump pipe replacement maintenance aspect
troublesomer, maintains 周期?uses the electrochemistry principle
the instrument to have to be short, maintains the work load to be big. From the influence aspect says to the environment, the potassium
bichromate dispels - the reduction method (or luminosity law, or
coulomb titrimetric method) has the chromium, the mercury two
contamination concern, the waste liquid needs special processing.
But UV counts the law and the electrification research method
(including coulomb titrimetric method) does not have this kind of
problem. 3.4 permanganates indexs analyzer The permanganate index on-line autoanalyzer main
technical principle has three kinds: (1) permanganate oxidation
- chemistry measures the mensuration; (2) permanganate oxidation
- electric current/potentiometric titration; (3) UV counts the
law (and the on-line COD meter similar). Says from the principle, the method (1) and the method (2) and
the non- essential difference (only is end point instruction way
difference), in country and so on Europe and America and Japan is the
legal method, with the our country standard method also is consistent.
(3) uses in the method to attribute the water quality
permanganate index the method, obtained a more widespread application
in Japan, but not yet promotes the application in our country, also
has not obtained the administrative hosts and visitors department the
approval. Says from the analysis performance, the present permanganate
index on-line autoanalyzer has been able to satisfy the surface water
on-line automatic monitor the need. Moreover, compares with the
color and chemistry method instrument, uses the reduction method
instrument analysis cycle generally longer somewhat (2h), former
generally is 15~60min. Says from the instrument structure, two kind of instruments
structures quite are complex. 3.5 总有机碳 (TOC) analyzer The TOC autoanalyzer in country's the and so on Europe
and America, Japan and Australia application is more widespread, its
main technical principle has four kinds: (1) (catalysis) burns
the oxidized - non- disperser infrared luminosity law (the NDIR law);
(2) the UV catalysis - peroxydisulfate oxidizes the -NDIR law;
(3) UV- peroxydisulfate oxidation - ion choice electrode law
(ISE) law; (4) heats up - the peroxydisulfate to oxidize the
-NDIR law; (5) the UV-TOC analysis counts the law. Says from the principle, only then shows (1) to approach the GB
code method, but the method (2) the ~ method (4) in country and so on
Europe and America also is the legal method. (5) uses in the
method to attribute water quality TOC, although obtained a more
widespread application in Japan, but not yet obtains the
administration department responsible for the work in the European and
American various countries the approval. Says from the analysis performance, the present on-line TOC
meter definitely can satisfy the source of pollution on-line automatic
monitor the need, and because its examination limits lowly, applies to
the surface water automatic monitor also is feasible. Moreover,
the on-line TOC meter analysis cycle is generally shorter (3~10min). Says from the instrument structure, removes the unit besides the
increase inorganic carbon, each kind of on-line TOC meter structure
on-line COD meter is generally simpler than some. 3.6 ammonia nitrogen and total nitrogen analyzer The ammonia nitrogen on-line autoanalyzer technical
principle mainly has three kinds: (1) ammonia sensitive
electrode 电位?(PH electrode law); (2) spectrophotometric
method; (3) Fourier transforms the spectrographic methods.
The on-line ammonia nitrogen meter and so on needs continuously
and the interrupted survey way, after passes through the on-line
filtrating equipment, the water sample measured the definite value
relative deviation is bigger. The total nitrogen on-line automatic minute meter main technical
principle has two kinds: (1) the peroxydisulfate dispels - the
luminosity law; (2) airtight burning oxidation - 化学发光
analytic method. 3.7 phosphate and total phosphorus analyzer (Reactivity) the phosphate autoanalyzer main technology
principle is the luminosity law. The total phosphorus on-line
autoanalyzer main technical principle includes: (1) the
peroxydisulfate dispels - the luminosity law; (2) the
ultraviolet ray illumination - molybdenum catalysis heats up dispels,
FLA- luminosity law. Says from the principle, the peroxydisulfate dispels - the
luminosity law is the on-line total nitrogen and the total phosphorus
meter host chooses the method, also is the various countries' legal
method. Heats up based on the airtight burning oxidation -
化学发光 analytic method on-line total nitrogen meter as well as
based on the ultraviolet ray illumination - molybdenum catalysis
dispels, the FIA- luminosity method on-line total phosphorus meter
mainly limits to Japan. Former is one of methods which the
Japanese industry specification association (JIS) approves. Says from the analysis performance, the present on-line total
nitrogen, the total phosphorus meter have been able to satisfy the
source of pollution and the surface water automatic monitor need, but
the sensitivity still with difficulty satisfied appraises kind of, two
kind of surface waters (standard value respectively was 0.04mg/l and
0.02mg/l) the water quality need. Moreover, uses 化学发光
the law, FIA- luminosity method instrument analysis cycle generally
shorter somewhat (10~30min), former generally is 30~60min. Says from the instrument structure, uses nonluminous method or
the FIA- luminosity method on-line total nitrogen, total phosphorus
meter structure simpler somewhat. 3.8 other analyze the instrument in the line segment TOD autoanalyzer: Technical principle generally
for burning oxidation - electrode law. Oil class autoanalyzer: The technical principle generally
is the fluorescence luminosity law. Phenol class autoanalyzer: The technical principle
generally is a color method. UV autoanalyzer: The technical principle is the color
method (254nm). Has simply, is quick, the price low
characteristic. Is ill to the surface water automatic on-line
monitor, overseas generally is uses in the source of pollution
automatic monitor, and frequently passes through the conversion to
express COD, the TOC value. Using the prerequisite is the water
quality is stabler, has the more definite linear correlation relations
in the UV absorption signal and COD or between the TOC value. Nitrate and cyanide autoanalyzer: The technical principle
mainly has: (1) ion choice electrode law; (2) luminosity
law. Fluoride and chloride autoanalyzer: Technical principle
generally for ion choice electrode law. |